CCD Laser Micrometer

VG series

This model has been discontinued.
Compliance with the certification standard is ensured as of the time of shipment from our company.

Recommended Replaceable Products: Multi-Purpose CCD Laser Micrometer - IG series

Amplifier Unit VG-301

VG-301 - Amplifier Unit

*Please note that accessories depicted in the image are for illustrative purposes only and may not be included with the product.

Specifications

Model

VG-301

Light source

Type

Visible red semiconductor laser

Maximum output

38 µW

Wavelength

670 nm

FDA (CDRH) 21CFR Part
1040.10

Class II Laser Product, Output : 38 µW

IEC 60825-1

Class 1 Laser Product, Output : 0.2 mW

Measuring area

35 mm

Setting distance

0 to 300 mm

Linearity

±0.1 % of F.S. (F.S. = 30 mm)*1*2

Minimumdisplayable unit

5 µm

Minimum detectable object

0.5 mm dia. opaque material*3

Repeatability

5 µm*4*5

Interface

RS-232C

Monitor output

Output voltage

±7 V

Impedance

100 Ω

Response time

3.9 ms (number of averaging measurements : 1)

Controloutput

Control

HI, GO, LO : NPN 100 mA, 40 V max.*6

Controloutput

NPN open collector : 100 mA max., (40 V max.) (one-shot output time : 1.3 ms)

Controloutput

Response time

3.9 ms (number of averaging measurements : 1)

Analogue output

Resolution (ripple)

2 mV (number of averaging measurements : 4096)

Resolution

5 µm (number of averaging measurements : 16)
Light-receiving element/CCD 5,000 bit, Scan time : 1.3 ms*7

Self-timing

NPN open collector : 100 mA max., (40 V max.) (one-shot output time : 1.3 ms)*6

Stability output

NPN open collector : 100 mA max., (40 V max.) (N.C.)*6

Response time

3.9 ms (number of averaging measurements : 1)

Digital I/O

RS-232C

Key input

10 keys

Control input

- Upper/ Lower limit select input
- Auto-zero input
- Auto-gain input
- Reset input
- Laser remote input voltage contact signals
- Hold input

Number of CCD pixels

Scanning time

Temperature
fluctuation

0.005 % of F.S./°C (F.S. = 30 mm)

Rating

Power voltage

24 VDC ±10 %, Ripple (P-P) 10 % or less

Current consumption

330 mA or less

Environmental resistance

Ambient light

Incandescent lamp: 1,000 lux or less, fluorescent lamp: 8,000 lux or less*8

Ambient temperature

0 to +50 °C

Relative humidity

35 to 85 % RH (No condensation)

Vibration resistance

10 to 55 Hz, Double amplitude 1.5 mm, 2 hours in each of the X, Y, and Z directions

Material

Polycarbonate

Weight

Approx. 800 g (including cable)

*1 F.S. (= 30 mm) indicates the range resulting from 2.5 mm subtraction from each end of the 35 mm measuring range.
*2 Typical data obtained when the T-to-R distance was 100 mm, the receiver-to-target distance was 50 mm and a knife edge was used to interrupt the laser beam.
*3 Typical data obtained when the T-to-R distance was 100 mm and the receiver-to-target distance was 50 mm .
*4 When the number of averaging measurements is 16
*5 This is a typical example when the distance between the transmitter and receiver has been set to 100 mm and a knife edge has been used to block light at a position that is 50 mm from the receiver. F.S. = 30 mm displays the measurement accuracy range.
*6 NPN output can easily be converted into PNP output by connecting the optional OP-5148 PNP output converter.
*7 Typical data obtained when the T-to-R distance was 100 mm, the receiver-to-target distance was 50 mm and a knife edge was used to interrupt the
laser beam.
*8 Typical data obtained when the T-to-R distance was 100 mm 3.94", the receiver-to-target distance was 50 mm 1.97" and a knife edge was used to interrupt the laser beam.

Data Sheet (PDF) Other Models