Specifications
Model  | LT-9011M  | |||
Type  | High-accuracy  | |||
Measurement range  | ±0.3 mm  | |||
Reference distance  | 6 mm  | |||
Light source  | Type  | Red semiconductor laser  | ||
Wavelength  | 655 nm  | |||
Output  | 170 µW (IEC 60825-1) /3.0 µW (FDA (CDRH) Part 1040.10)  | |||
Laser Class  | Class IIa Laser Product (FDA (CDRH) Part 1040.10)  | |||
Spot diameter  | Approx. ø2 µm  | |||
Scan width/interval  | 0 to 1,100 µm (6 steps) /1 to 10 µm (4 steps)  | |||
Resolution  | 0.3 µm*1  | |||
Linearity  | ±0.5 % of F.S.*1  | |||
Sampling cycle  | 640 µs to 356 ms (14 steps)*2  | |||
Display  | Minimumdisplayable unit  | 0.1 µm*3, 1 µm2, 0.01 °  | ||
Display range  | ±9,999.99 µm, ±999,999 µm2, 9,999.99 °  | |||
Microscope function  | Available  | |||
Display cycle  | 10 times/sec.*4  | |||
Terminal block  | Analogue output  | ±10 V x 2 outputs, output impedance: 100 Ω  | ||
Timing input  | Non-voltage input  | |||
Reset input  | ||||
Auto-zero input  | ||||
Monitor dedicated power supply  | 24 VDC*5  | |||
Control I/O  | Judgment mode  | 3-step limits output  | For OUT1 and OUT2, and NPN open collector output*6  | |
Binary mode  | Binary output  | Measured data output (21 bits), OUT1/OUT2/PROFILE selectable  | ||
Strobe output  | NPN open collector output*6  | |||
Binary selector output  | ||||
Binary selection input  | Non-voltage input*6  | |||
Stability output  | NPN open collector output  | |||
Laser remote input  | Non-voltage input  | |||
Program switching input  | Non-voltage input x 3 inputs  | |||
RS-232C interface  | Measured data output and control I/O (Selectable up to baud rate 115200 bps.)  | |||
Video output  | NTSC compliant (PIN connector)  | |||
Microscope function  | Function presence/absence  | Compatible  | ||
Field of view  | 1.3 mm x 1.05 mm  | |||
Type  | Available  | |||
Illumination light source  | Infrared LED  | |||
Main functions  | Displacement mode  | Distance measurement, Transparent objectthickness measurement, Angle measurement,Relative angle measurement, Surface selection,Dark-out, Mask, Trend graph display, and Scanwidth/interval change*7  | ||
Profile mode  | Area selection (Average, Maximum,  | |||
Common  | Light intensity accumulation, Microscope (LT-9501H, LT-9501HSO (5654) only),  | |||
Temperature characteristics (20 to 30°C)  | ±0.5% of F.S.  | |||
Rating  | Power voltage  | 100 to 240 VAC ±10 %, 50/60 Hz  | ||
Power consumption  | 110VA or less  | |||
Environmental resistance  | Pollution degree  | 2  | ||
Overvoltage category  | II  | |||
Ambient light  | Incandescent lamp/fluorescent lamp: 10,000 lux max.  | |||
Ambient temperature  | 0 to +35 °C  | |||
Relative humidity  | 35 to 85 % RH (No condensation)  | |||
Weight  | Approx. 400 g  | |||
*1 The value when the measurement target is a mirrored surface object measured in displacement mode, scan width/interval 120 µm/2 µm, and 8-times average  | ||||